
7
7-8 z.one ultra and z.one ultra sp OPERATOR MANUAL
To Enable/Disable a Factory or User Preset in Menu
NOTE: Factory Presets are identified by a Zonare Z inside the bubble to the left of
their name in the main Presets menu. Customized User Presets can be identified
by an empty bubble to the left of their name.
1. Press Exam Type button.
2. Ensure that the desired Exam Type is currently selected.
NOTE: If not, go to the Presets tab and select desired Exam Type prior to beginning
the Enable/Disable steps below.
3. Select the Preset Management option.
4. Highlight and select Enable/Disable option. The Enable/Disable Presets
screen displays.
Figure 7-4. Enable/Disable Presets Screen
5. Uncheck the box to the left of any Factory or User Preset that you wish to
REMOVE (disable) from the Presets menu. (In
Figure 7-4, no User Presets
have been configured.)
6. Highlight and select Apply to activate the new list. Only the checked Presets
will display on the Preset list.
NOTE: You can enable/disable Presets for each transducer by selecting the
transducer and Exam Type via the drop-down menus.
To Configure Softkey Display of Available Presets
SmartCart/SmartCart sp workstations displays up to 6 Exam/Preset softkeys for
the active transducer. You can customize the display of available Exam/Preset
softkeys as described below.
1. Go to Exam Type button | Preset Mgmt | Customize.
2. The Preset Customization Configuration screen displays:
Komentáře k této Příručce