7-52 z.one ultra and z.one ultra sp OPERATOR MANUAL
Figure 7-21. PW Autotrace Icon - Doppler Data Field
3. Select the desired measurement (e.g., RI) and press Set to initiate the Trace.
4. To select a different measurement or calc, press Measure or Calc button to
display menu. Make your selection and press Set.
To Deactivate Auto-Dop Trace
Three methods:
Press Function/Mode key assigned to Autotrace (see page 9-46).
Double-click Measure button.
Display Measure or Calc menu (Figure 7-20) and click Autotrace option.
Continuous Wave (CW) Doppler Imaging
Continuous Wave (CW) Doppler imaging is available only on SmartCart systems
equipped with the Echocardiology Option and the P4-1c transducer.
Auxiliary CW Doppler also operates only on SmartCart systems equipped with the
Echocardiology Option.
NOTE: See Figure 10-18 on page 10-15 for a description of the Doppler Data
Field. See also Chapter 10 Echocardiography Option for information about CW
and Aux CW and cardiac imaging.
To Activate CW Doppler Imaging
1. Connect P4-1c transducer to MTP.
2. Acquire B-Mode image; then press D button to activate PW.
3. Turn right-most softkey to select CW (options are CW, PW, and TDI [tissue
doppler imaging]); then press button to activate CW Doppler imaging.
Figure 7-22. Doppler Mode Softkey (far right on softkey panel)
Indicates you are in
Autotrace mode.
TDI
PW
CW
Turn to select desired Doppler mode
Then press to activate selection