9-18 z.one ultra and z.one ultra sp OPERATOR MANUAL
NOTE: Any gap in the outline between first and second Trace calipers will be
automatically closed by a straight connecting line when second Trace caliper
position is fixed.
8. To undo any portions of the trace:
9. To activate additional Trace circumference/area measurements (up to a
maximum of two), press Select button to invoke another Trace measurement
instance.
10. Press Measure button to remove the Trace tools and exit Generic
Measurements.
Volume (cm
3
) To Measure Volume
1. Acquire long axis image first.
2. Freeze 2D image; then press Measure button.
3. Use Trackball and Set button to select Volume.
4. Use Trackball to position first caliper for caliper set 1. Press Set.
5. Use Trackball to position next caliper.
6. Press Enter (Enter/Select on sp).
7. Repeat Steps 4 and 5 for caliper set 2. Press Store button.
8. Unfreeze image. Acquire short axis view.
9. Freeze image; then press Measure.
10. Use Trackball to position first caliper for caliper set 3. Press Set.
11. Use Trackball to position next caliper. Press Store. Volume calculation
displays on screen.
M-Mode
Depth In M-Mode, Depth measures the vertical distance from the top of the strip
(transducer face) to the position of the cursor/caliper in the M-Mode strip. The
DDB displays the Depth measurement result and continuously updates the result
as the user positions the cursor/caliper within the image.
SmartCart Press Delete button; each press removes last
small dot
SmartCart sp
Press Back/Delete button; each press removes
last small dot
Scan Engine Press Back button; each press removes last small
dot