6-10 z.one ultra and z.one ultra sp OPERATOR MANUAL
Tip Deflection Control
The deflection control wheels found on the handle of the transducer control
deflection of the transducer tip.
Figure 6-7. Tip Deflection Control Wheels
WARNING: Check if the maximum deflection of the tip is 90° to 120° upward,
60° to 90° downwards and 30° to 45° left/right. If the deflection shows an
unwanted amount of free play or exceeds the maximal deflection (displaying a
sharp “U- turn”), do not use the transducer. Contact a
ZONARE service
representative to report the problem.
Being aware of the tip deflection movement minimizes the risk of “folding
over” or “U-turning” of the transducer in the patient. If you believe the
transducer has folded over in the patient, make sure the wheels are in the
unlock position and gently try manipulating the transducer. Try advancing the
transducer into the stomach and attempt to straighten the folded over
transducer and withdraw from the patient.
Friction Lock Operation
To protect the patient and the transducer, unlock the deflection controls when
inserting or withdrawing the transducer. The friction lock is used for locking either
the Anterior/Posterior or Right/Left
TEE tip deflection.
Figure 6-8. Friction Lock Operation - Right/Left
Anterior/Posterior Wheel
Right/Left Wheel
White
Unlock
Lock
Red
UP / DOWN
Right/Left