14
14-18 z.one ultra and z.one ultra sp OPERATOR MANUAL
3. Press the Power button on the Scan Engine, and then immediately make note
of the battery charge LED indicators (located on the lower left side of the
Scan
Engine
; see Figure 4-4 on page 4-5). All four battery LED indicators are lit
when the battery is fully charged. The four LED indicators are illuminated in
25% increments:
• 4 LEDs = 76-100% charge
• 3 LEDs = 51-75% charge
• 2 LEDs = 26-50% charge
• 1 LED = 0-25% charge
4. If you are certain the battery is fully charged (as described above) and the
Scan Engine does not perform accordingly, replace the battery or contact
ZONARE Technical Support for assistance. There are no user remedies for
battery insufficiency problems other than maintaining a full charge.
Battery
Charging
(Scan Engine)
For optimum performance, the battery packs should be maintained in a charged
state. Battery charging occurs when the
SmartCart/SmartCart sp is connected to
an active power receptacle, the battery pack is inserted into the
Scan Engine, and
the
Scan Engine is stored in a docked condition on the SmartCart/SmartCart sp.
Remote charging and recalibrating (resetting of the charge fuel gauge) of the
battery pack is done using the ZONARE 2-Bay Battery Charger unit (Part # Z311:
for 110V or #Z312: for 220V).
Battery
Maintenance
(Scan Engine)
The battery is a spill-proof maintenance-free sealed battery. There are no user
actions for battery maintenance other than maintaining a charged state.
Typical battery life is one to two years when the battery is maintained in a charged
state. Leaving the battery in an uncharged state reduces its life. Using the Zonare
2-Bay Battery Charger’s “Recalibrate” (recondition) cycle on a scheduled basis
will help maintain optimum efficiency and performance of the batteries.
CAUTION: If the SmartCart/SmartCart sp will be stored in a location where power
cannot be supplied (to maintain the battery in a charged state) for more than five
days, the battery should be removed from the
Scan Engine to help prevent
potentially hazardous conditions as a result of the battery deteriorating.
WARNING: To protect the Scan Engine battery from potential thermal damage,
the System monitors the temperature of the battery at all times. If the battery is
detected as exceeding the maximum safe operating temperature, a warning
message will appear on the
SmartCart/SmartCart sp display.
NOTE: The Operator should immediately complete all active operations in
preparation for the System automatically initiating a power-down sequence.
Internal
Storage Media
Maintenance
The Scan Engine internal storage media should not be used for long-term storage
of studies and patient information. Studies should be deleted from the storage
media on a regular basis (daily or weekly) as part of normal User maintenance on
the
Scan Engine.