6-16 z.one ultra and z.one ultra sp OPERATOR MANUAL
Cleaning Wipe Method
1. Disconnect the transducer from the Cart. If present, remove the transducer
cover/sheath and discard.
2. Use a soft cloth dampened with a mild soap (Liquinox) and water solution to
remove particulate, ultrasound coupling gel, or body fluids.
3. Next, ZONARE recommends the use of a disinfectant solution to wipe the
strain relief, control handle, deflection wheel, push buttons, cable and
transducer connector. Avoiding touching the internal pins with any cleaner.
Follow the solution manufacturer's instructions for proper cleaning.
NOTE: A list of approved disinfecting solutions can be found later in this document
4. Air-dry (or towel dry) the transducer with a soft, clean cloth.
5. Examine the transducer for damage, such as cracks, splitting, holes, or fluid
leaks. If damage is evident, discontinue use of the transducer and contact
ZONARE Technical Support.
NOTE: The multiplane TEE transducer is a precision instrument that must be
handled with care. It may be damaged when dropped or abused. In particular, do
not allow the distal transducer tip to come in contact with a sharp object. Do not
touch the ultrasonic window in the tip unnecessarily. Never exert force onto the
window.
WARNING: Keep the control handle and transducer connector out of any cleaning
or disinfection solutions. The control handle and cable may be cleaned with a
damp cloth, but only the distal end of the transducer up to the 100-cm marker on
the shaft may be placed into a disinfection solution.
Disinfecting Immersion Method
CAUTION: During immersion disinfection, never immerse transducers longer than
45 minutes. Damage may occur to the transducer housing and/or components if
disinfection times exceed these recommended limits.
CAUTION: Using a nonrecommended cleaning or disinfectant solution, incorrect
solution strength, or immersing the transducer deeper or longer than indicated
can damage the transducer.
1. Follow the disinfectant manufacturer's recommendations for solution
preparation, immersion time and rinsing.
2. Place the cleaned and dried transducer surface in contact with the disinfection
solution. Take care that the flexible shaft is immersed up to no more than the
100-cm mark (see
Figure 6-6).
3. Rinse the transducer tip and shaft with water and dry prior to storage.
WARNING: Do not use other disinfection methods like iodine, steam, heat, or
ethylene oxide.
WARNING: The use of a biteguard is mandatory. Failure to use the biteguard may
result in damage to the transducer, which could result in a safety hazard. Damage
to the transducer due to biting is not covered by the transducer's warranty.
TEE Transducer Cleaning and Disinfecting Solutions