4
z.one ultra and z.one ultra sp OPERATOR MANUAL 4-3
•The Scan Engine's internal battery is automatically recharged while the
Scan Engine is docked.
■
When the Scan Engine is undocked:
• Images are displayed on the Scan Engine’s local LCD Display (see “Scan
Engine LCD Display” on page 5-1).
• User operations are performed on the Scan Engine’s local Control Panel
(see
“Scan Engine Control Panel” on page 5-3).
•The Scan Engine is powered by an internal rechargeable battery (or the
optional AC adapter; see
page 4-4).
Internal
Archive
Storage
The Scan Engine base model comes standard with 512 MB of internal storage
capacity. Storage capacity of 2 GB is available as an option.
NOTE: The Scan Engine internal hard drive (SSD/C:) is not intended to be used as
a long-term archive. All patient exams and system presets should be backed up
regularly. See
page 11-21 and page 14-6 for more information.
NOTE: You can also save exams directly to the SmartCart/SmartCart sp’s internal
hard drive (CART HD); see
Table 11-1 on page 11-3 and page 11-4.
Scan Engine
(Undocked)
To Power ON the Scan Engine
1. Make sure the Scan Engine battery is charged. See “Checking Scan Engine
Battery Charge” on page 4-5.
2. Press the power switch on the Scan Engine (Figure 4-2). After pressing the
switch, it remains illuminated for 10 seconds and then turns off.
NOTE: The system is still ON even though the light is out.
3. The ZONARE startup screen displays while the system initializes. When
initializing is complete (takes about 20 seconds), the System is ready for 2D
(B-Mode) imaging.
• If the screen is blank or not the default display, check the transducer
connections and reconnect the transducer.
• If a normal display is still not present, connect a different transducer.
• If the normal display is still not present, power the system OFF and call
Technical Support (see
“ZONARE Contact Information” on page 15-1).
To Power OFF the Scan Engine
1. Press and release the Scan Engine power switch. The Scan Engine takes
about 10 seconds to power
OFF.
NOTE: Do not press and hold down the power switch. Instead, just press and
release.